This screen allows you to add Schedule Overrides such as emergency closures or holidays in which a facility is closed. Typically, an agency will run three consecutive 10mg tests each morning to confirm the pump is producing exactly 10mg. Additionally, this screen is required to empty Methadone out of the pump at the end of each day and to insert distilled water into the pump for cleaning. In the morning, this process is reversed by clearing the distilled water out of the pump and placing Methadone on the pump, prime the pump to prepare for dispensing.
To access this screen, go to Administration 🡪 Pump Calibration.
Steps for Testing the Connection and Calibrating the Pump
- Select the pump from the drop-down.
- Select the appropriate COM port.
-
Refresh Pump – Click the Refresh Pump to test the connection.
- You will see Initializing Pump… in the header while the connection is being tested.
- If the connection is unsuccessful, you will see something like “The port ‘COM3’ does not exist.”
Description of Options
- Door Alarm – Sends command to the pump to enable or disable the door alarm.
- Liquid Eye – Sends command to the pump to enable or disable the liquid eye.
- Mode – Sends command to the pump to prime or dispense.
- Pump Direction – Indicates the direction in which the pump is pumping (Into or out of the bottle).
- Clear Hose – Sends command to pump (Pumps in reverse to empty liquid back into the bottle).
- Prime – Sends command to pump (Primes the pump).
- mL – Amount of medication to be dispensed from the pump in mL once you hit the “Begin” button.
- Begin – Sends command to pump to dispense the amount indicated in the mL field (At left).
-
Commands – Displays the commands that are being sent to and received from the connected pump.
Clear Hose
Clicking Clear Hose will automatically dispense in reverse, sending any liquid in the hose back into the bottle. No buttons need to be click other than the Clear Hose button for it to run.
Prime
Clicking Prime will prime the pump, dispensing about 60 mL (per manufacturer, this amount can vary) of liquid. No buttons need to be click other than the Prime button for it to run.
Dispensing Liquid
From the Pump Calibration screen, the end-user can dispense liquid from the bottle.
- Set Mode to Dispense.
- NOTE: If it is set to Prime do not drag dot, just click on the dot and it will move.